The DP30K-0002HV is suitable, for example, for electrostatic test setups, testing high-voltage components under load and current-limited charging tests on capacitive test loads. The output provides 0kV ... 30kV over a current range of 0mA ... 2mA with a rated power of 60 W. You can adjust voltage and current at the front panel or integrate the unit into a test system through the standard analog I/O or optional RS485 interface.

Precision and output parameters
Voltage and current setpoints can be adjusted independently. CV/CC regulation switches automatically between constant-voltage and constant-current operation. Precision rotary controls and separate LED displays for voltage and current support local adjustment and monitoring.
| Parameter | Voltage | Current |
|---|---|---|
| Setting and display resolution | 10V | 0,001mA |
| Setting accuracy | ±0,5% FS ±2 dig | ±0,5% FS ±2 dig |
| Display accuracy | ±0,5% FS ±2 dig | ±0,5% FS ±2 dig |
| Line and load regulation | Each ≤1% FS | Each ≤1% FS |
| Ripple | ≤1% FS RMS | ≤1% FS RMS |
| Temperature coefficient | 100ppm/°C | 100ppm/°C |
| Output Configuration | Grounded, with positive or negative polarity as ordered | |
FS denotes full scale and dig the last display digit.
Test procedures and interfaces
In the standard analog configuration, separate 0V ... 10V signals set voltage and current commands. Corresponding measured values are also available as 0V ... 10V signals. Additional status outputs indicate CV/CC operation, faults and the state of the high-voltage output.
RS485 with Modbus RTU can be ordered for digital integration. The test-bench controller or a controller/computer can then set commands and read voltage and current measurements. Command response time is 50 ms. The RS485 version replaces the analog I/O, including its status outputs and auxiliary reference. The standard interlock loop is available in both versions.
| Configuration | Control and feedback |
|---|---|
| Analog, standard | 0V ... 10V for U/I setpoints and measured values, plus status outputs |
| RS485, optional | Modbus RTU for digital setpoints and measurement queries, with Windows control software |
The Windows control software supplied with the optional RS485 interface supports operation from a PC. The documented Modbus RTU protocol is also available for your own test programs.

Output configuration and installation
The output voltage is configured as positive or negative with respect to earth. The required polarity is specified when ordering.
The DP30K-0002HV measures 323 × 110 × 367 mm (W × H × D) and weighs 14 kg. Its mains input is single-phase, 207V ... 253V AC at 47Hz ... 63Hz. The fans are temperature-controlled.
Similar models
The following DP-HV models offer the same maximum output voltage, the same maximum output current or the same rated power as the DP30K-0002HV.
Same output voltage: 30 kV
| Model | Output Current | Rated power |
|---|---|---|
| DP30K-0001HV | 0 ... 1 mA | 30 W |
| DP30K-0005HV | 0 ... 5 mA | 150 W |
| DP30K-001HV | 0 ... 10 mA | 300 W |
| DP30K-002HV | 0 ... 20 mA | 600 W |
| DP30K-003HV | 0 ... 30 mA | 900 W |
| DP30K-004HV | 0 ... 40 mA | 1.2 kW |
| DP30K-005HV | 0 ... 50 mA | 1.5 kW |
| DP30K-006HV | 0 ... 60 mA | 1.8 kW |
| DP30K-008HV | 0 ... 80 mA | 2.4 kW |
| DP30K-01HV | 0 ... 100 mA | 3 kW |
| DP30K-012HV | 0 ... 120 mA | 3.6 kW |
Same output current: 2 mA
| Model | Output Voltage | Rated power |
|---|---|---|
| DP15K-0002HV | 0 ... 15 kV | 30 W |
| DP20K-0002HV | 0 ... 20 kV | 40 W |
| DP50K-0002HV | 0 ... 50 kV | 100 W |
| DP80K-0002HV | 0 ... 80 kV | 160 W |
| DP120K-0002HV | 0 ... 120 kV | 240 W |
| DP150K-0002HV | 0 ... 150 kV | 300 W |
Same rated power: 60 W
| Model | Output Voltage | Output Current |
|---|---|---|
| DP15K-0004HV | 0 ... 15 kV | 0 ... 4 mA |
| DP20K-0003HV | 0 ... 20 kV | 0 ... 3 mA |
| DP40K-00015HV | 0 ... 40 kV | 0 ... 1,5 mA |
| DP60K-0001HV | 0 ... 60 kV | 0 ... 1 mA |
| DP120K-00005HV | 0 ... 120 kV | 0 ... 0,5 mA |
| DP130K-000046HV | 0 ... 130 kV | 0 ... 0,46 mA |