Bench power supply accuracy: What the percentage means at the test point

A bench power supply displays 5,000 V, and its datasheet specifies an accuracy of 0,1 %. That may seem sufficient for selection. But before interpreting it as an allowable deviation of 5 mV, one essential detail is missing: what is the percentage based on?

For a test bench with tight voltage tolerances, this question matters more than another decimal place. Only the complete specification reveals the deviation specified at the intended test point. You must also distinguish between setting a voltage and measuring it with the power supply.

The same 0,1 % can mean different things

Consider two explicitly hypothetical specifications for a power supply's voltage measurement. They are calculation examples and do not describe a particular model.

Specification A is ±(0,1 % of full scale + 2 mV). Full scale, explicitly assumed to be 60 V here, is the reference quantity. This gives:

0,001 × 60 V + 0,002 V = 0,062 V = 62 mV.

In this example, the absolute error limit is the same at 5 V, 15 V and 50 V. Relative to the displayed voltage, however, its significance varies considerably.

Specification B is instead ±(0,1 % of reading + 2 mV). Here, the percentage term grows with the measured value. At 5 V, the calculation gives only 0,001 × 5 V + 0,002 V = 7 mV.

Displayed value A: Error limit at 60 V full scale A: Relative to displayed value B: Error limit based on reading
5 V ±62 mV ±1,24 % ±7 mV
15 V ±62 mV approx. ±0,413 % ±17 mV
50 V ±62 mV ±0,124 % ±52 mV
Comparison of two hypothetical error limits at 5, 15 and 50 V: 0,1 percent of a 60 V full scale plus 2 mV gives 62 mV at every point. A reading-based 0,1 percent plus 2 mV gives 7, 17 and 52 mV.
Hypothetical error limits: the same percentage, different reference quantities. This is not a measurement curve of a real device.

The difference comes from the formula. You cannot turn a full-scale specification into a reading-based specification by substituting your test voltage for 60 V. The actual datasheet must also establish which range is meant by Full Scale. Where measurement ranges are switchable, the selected range may be the relevant one.

For procurement, it is therefore worth calculating the limit at the lowest critical voltage. A source whose measurement function appears adequate for a 50 V test has not thereby been selected for evaluating tight deviations on a 5 V circuit.

More display digits do not reduce the limit

A display with 1 mV increments can reveal small changes. That does not promise an accuracy of ±1 mV. Resolution describes distinguishable increments, while accuracy concerns deviation from the correct value. Rohde & Schwarz also explains this distinction for measurement and monitoring with DC power supplies.

In hypothetical case A, displaying 5,000 V therefore does not make the measurement accurate to a millivolt. The assumed limit remains ±62 mV. Conversely, this limit does not mean that every measurement actually deviates by 62 mV. It describes the allowable bounds under the assumed specification conditions.

Setting and measuring are different tasks

Setting accuracy describes how close the actual output is to the programmed setpoint. Measurement or readback accuracy describes how accurately the device measures its output. Different resolutions and accuracies may apply to these functions. Rohde & Schwarz explicitly distinguishes them.

Check the datasheet row that corresponds to your task. An accurately set value does not establish measurement accuracy. Leads and contacts between the output and the device under test can also cause voltage drops. The defined measurement point matters for the assessment. Use any available sense functions according to the device instructions.

Conditions are part of the specification

A complete selection also includes the conditions under which the manufacturer specifies accuracy. Check the temperature range, any required warm-up time, the measurement range and the stated interval since calibration. Clarify missing conditions rather than assuming unlimited validity.

Then compare the same function under comparable conditions. One device's setting specification and another's measurement specification answer different questions. A typical value is not automatically equivalent to a guaranteed limit either. Your test decision requires the specified property at the relevant operating point.

From the datasheet to the test setup

An error limit is neither a standard uncertainty nor an automatically valid 95 % interval. An uncertainty assessment requires the available information and its underlying assumptions to be consistent. Manufacturer specifications are one possible input alongside calibration data and other information, as NIST describes for evaluating uncertainty contributions.

For your setup, the power supply specification is therefore just one contribution. First define which quantity is to be assessed, at which point, and what deviation the test permits. You can then determine the requirements for the source, measuring instrument and connections. Contributions already accounted for should not be added again as a blanket allowance.

The DC bench power supplies in the Laborance catalogue are a starting point for selection. For a focused enquiry to service@dsc-electronics.de, provide the critical voltage values, allowable deviation in millivolts and measurement point. This lets us check the relevant specification before an easy-to-read display is mistaken for an accuracy guarantee.