The DF-HV30KV-67MA-40KHZ is suitable, for example, for high-frequency excitation of dielectric test objects and for experimental setups investigating materials and discharges. It combines adjustable sinusoidal high voltage up to 30 kV peak with a fixed output frequency of 40 kHz. Voltage can be adjusted locally to suit the test object.

Output parameters and regulation
Voltage and current limit can be adjusted independently. When the load current reaches the set limit, the regulator reduces the internal DC-link voltage and therefore the AC output voltage.
With an appropriate setting, this voltage reduction can limit or extinguish arcs. Integrated OVP limits the peak voltage to the specified upper limit.
| Parameter | DF-HV30KV-67MA-40KHZ |
|---|---|
| Output Voltage | 0-30 kV peak |
| Maximum output current | 67 mA |
| Peak power | 2 kW |
| Output Frequency | 40 kHz, fixed |
Test procedures and local controls
Voltage and current limit are adjusted using ten-turn controls on the front panel. This model has a fixed output frequency of 40 kHz.
| Display | Configuration |
|---|---|
| Voltage | Four-digit LED display, 0,01 kV resolution |
| Current | Four-digit LED display |
| Frequency | Four-digit LED display, 0,01 kHz resolution |

Output configuration and installation
The resonant inverter uses soft switching. The unit is factory-matched at the ordered frequency to the intended load circuit, including its cables and electrodes.
This configuration consists of a control module and an HV module. Each module measures 439 × 133 × 436 mm (W × H × D). The control module weighs approximately 12 kg and the HV module approximately 24 kg. The system is fan-cooled and operates from a single-phase 230 V ±10 % supply at 50/60 Hz. The operating environment is 0-50 °C and 10-80 % relative humidity.
Similar models
The following DF-HV models offer the same peak voltage, maximum output current or peak power. Frequency variants have either a fixed value or an adjustable range.
Same peak voltage: 30 kV
| Model | Peak U | Max. current | Peak P | Frequency |
|---|---|---|---|---|
| DF-HV30KV-30MA-5KHZ | 30 kV | 30 mA | 900 W | 5 kHz |
| DF-HV30KV-30MA-5KHZ-10KHZ | 30 kV | 30 mA | 900 W | 5-10 kHz |
| DF-HV30KV-30MA-10KHZ | 30 kV | 30 mA | 900 W | 10 kHz |
| DF-HV30KV-30MA-10KHZ-20KHZ | 30 kV | 30 mA | 900 W | 10-20 kHz |
| DF-HV30KV-30MA-10KHZ-30KHZ | 30 kV | 30 mA | 900 W | 10-30 kHz |
| DF-HV30KV-30MA-20KHZ | 30 kV | 30 mA | 900 W | 20 kHz |
| DF-HV30KV-30MA-20KHZ-30KHZ | 30 kV | 30 mA | 900 W | 20-30 kHz |
| DF-HV30KV-30MA-20KHZ-40KHZ | 30 kV | 30 mA | 900 W | 20-40 kHz |
| DF-HV30KV-30MA-20KHZ-50KHZ | 30 kV | 30 mA | 900 W | 20-50 kHz |
| DF-HV30KV-30MA-30KHZ | 30 kV | 30 mA | 900 W | 30 kHz |
| DF-HV30KV-30MA-30KHZ-40KHZ | 30 kV | 30 mA | 900 W | 30-40 kHz |
| DF-HV30KV-30MA-30KHZ-50KHZ | 30 kV | 30 mA | 900 W | 30-50 kHz |
| DF-HV30KV-30MA-40KHZ | 30 kV | 30 mA | 900 W | 40 kHz |
| DF-HV30KV-30MA-50KHZ | 30 kV | 30 mA | 900 W | 50 kHz |
| DF-HV30KV-30MA-50KHZ-100KHZ | 30 kV | 30 mA | 900 W | 50-100 kHz |
| DF-HV30KV-30MA-100KHZ | 30 kV | 30 mA | 900 W | 100 kHz |