The DF-HV30KV-30MA-20KHZ-30KHZ is suitable, for example, for high-frequency excitation of dielectric test objects and for experimental setups investigating materials and discharges. It combines adjustable sinusoidal high voltage up to 30 kV peak with a frequency range of 20 to 30 kHz. Voltage and frequency can be adjusted locally to suit the test object.

Output parameters and regulation
Voltage and current limit can be adjusted independently. When the load current reaches the set limit, the regulator reduces the internal DC-link voltage and therefore the AC output voltage.
With an appropriate setting, this voltage reduction can limit or extinguish arcs. Integrated OVP limits the peak voltage to the specified upper limit.
| Parameter | DF-HV30KV-30MA-20KHZ-30KHZ |
|---|---|
| Output Voltage | 0-30 kV peak |
| Maximum output current | 30 mA |
| Peak power | 900 W |
| Output Frequency | 20-30 kHz, adjustable |
| Frequency accuracy | ≤5 % of the centre frequency |
Test procedures and local controls
Voltage, current limit and frequency are adjusted using three ten-turn controls on the front panel. With dielectric loads, a change in frequency also changes the current drawn by the test system.
| Display | Configuration |
|---|---|
| Voltage | Four-digit LED display, 0,01 kV resolution |
| Current | Four-digit LED display |
| Frequency | Four-digit LED display, 0,01 kHz resolution |

Output configuration and installation
The resonant inverter uses soft switching. The unit is factory-matched within the ordered frequency band to the intended load circuit, including its cables and electrodes.
The unit weighs approximately 22 kg. It is fan-cooled and operates from a single-phase 230 V ±10 % supply at 50/60 Hz. The operating environment is 0-50 °C and 10-80 % relative humidity.
Similar models
The following DF-HV models offer the same peak voltage, maximum output current or peak power. The frequency versions differ by a fixed value or an adjustable range.
Same peak voltage: 30 kV
| Model | Max. current | Peak P | Frequency |
|---|---|---|---|
| DF-HV30KV-67MA-5KHZ | 67 mA | 2 kW | 5 kHz |
| DF-HV30KV-67MA-8KHZ | 67 mA | 2 kW | 8 kHz |
| DF-HV30KV-67MA-10KHZ | 67 mA | 2 kW | 10 kHz |
| DF-HV30KV-67MA-20KHZ | 67 mA | 2 kW | 20 kHz |
| DF-HV30KV-67MA-30KHZ | 67 mA | 2 kW | 30 kHz |
| DF-HV30KV-67MA-40KHZ | 67 mA | 2 kW | 40 kHz |
| DF-HV30KV-67MA-50KHZ | 67 mA | 2 kW | 50 kHz |
| DF-HV30KV-67MA-100KHZ | 67 mA | 2 kW | 100 kHz |
| DF-HV30KV-67MA-5KHZ-10KHZ | 67 mA | 2 kW | 5-10 kHz |
| DF-HV30KV-67MA-10KHZ-20KHZ | 67 mA | 2 kW | 10-20 kHz |
| DF-HV30KV-67MA-10KHZ-30KHZ | 67 mA | 2 kW | 10-30 kHz |
| DF-HV30KV-67MA-20KHZ-30KHZ | 67 mA | 2 kW | 20-30 kHz |
| DF-HV30KV-67MA-20KHZ-40KHZ | 67 mA | 2 kW | 20-40 kHz |
| DF-HV30KV-67MA-20KHZ-50KHZ | 67 mA | 2 kW | 20-50 kHz |
| DF-HV30KV-67MA-30KHZ-40KHZ | 67 mA | 2 kW | 30-40 kHz |
| DF-HV30KV-67MA-30KHZ-50KHZ | 67 mA | 2 kW | 30-50 kHz |
| DF-HV30KV-67MA-50KHZ-100KHZ | 67 mA | 2 kW | 50-100 kHz |