The DF-HV15KV-200MA-30KHZ-50KHZ is suitable, for example, for high-frequency excitation of dielectric test objects and for experimental setups investigating materials and discharges. It combines adjustable sinusoidal high voltage up to 15 kV peak with a frequency range of 30 to 50 kHz. Voltage and frequency can be adjusted locally to suit the test object.

Output parameters and regulation
Voltage and current limit can be adjusted independently. When the load current reaches the set limit, the regulator reduces the internal DC-link voltage and therefore the AC output voltage.
With an appropriate setting, this voltage reduction can limit or extinguish arcs. Integrated OVP limits the peak voltage to the specified upper limit.
| Parameter | DF-HV15KV-200MA-30KHZ-50KHZ |
|---|---|
| Output Voltage | 0kVp ... 15kVp |
| Maximum output current | 200 mA |
| Power | 0kW ... 3kW |
| Output Frequency | 30kHz ... 50kHz, einstellbar |
| Frequency accuracy | ≤5 % of the centre frequency |
Test procedures and local controls
Voltage, current limit and frequency are adjusted using three ten-turn controls on the front panel. With dielectric loads, a change in frequency also changes the current drawn by the test system.
| Display | Configuration |
|---|---|
| Voltage | Four-digit LED display, 0,01 kV resolution |
| Current | Four-digit LED display |
| Frequency | Four-digit LED display, 0,01 kHz resolution |

Output configuration and installation
The resonant inverter uses soft switching. The unit is factory-matched within the ordered frequency band to the intended load circuit, including its cables and electrodes.
Das Gerät misst 439 × 133 × 563 mm (B × H × T) und wiegt etwa 20 kg. Es wird einphasig mit 230 V ±10 % bei 50/60 Hz versorgt und ist lüftergekühlt. Die zulässige Betriebsumgebung liegt bei 0°C ... 50°C und 10% ... 80% relativer Luftfeuchte.
Similar models
The following DF-HV models offer the same maximum output current as the DF-HV15KV-200MA-30KHZ-50KHZ.
Same output current: 200 mA
| Model | Voltage | Power | Frequency |
|---|---|---|---|
| DF-HV20KV-200MA-1KHZ | 20 kVp | 4 kW | 1 kHz |
| DF-HV20KV-200MA-2KHZ | 20 kVp | 4 kW | 2 kHz |
| DF-HV20KV-200MA-2KHZ-8KHZ | 20 kVp | 4 kW | 2kHz ... 8kHz |
| DF-HV20KV-200MA-3KHZ | 20 kVp | 4 kW | 3 kHz |
| DF-HV20KV-200MA-4KHZ | 20 kVp | 4 kW | 4 kHz |
| DF-HV20KV-200MA-5KHZ | 20 kVp | 4 kW | 5 kHz |
| DF-HV20KV-200MA-5KHZ-10KHZ | 20 kVp | 4 kW | 5kHz ... 10kHz |
| DF-HV20KV-200MA-8KHZ | 20 kVp | 4 kW | 8 kHz |
| DF-HV20KV-200MA-10KHZ | 20 kVp | 4 kW | 10 kHz |
| DF-HV20KV-200MA-10KHZ-20KHZ | 20 kVp | 4 kW | 10kHz ... 20kHz |
| DF-HV20KV-200MA-10KHZ-30KHZ | 20 kVp | 4 kW | 10kHz ... 30kHz |
| DF-HV20KV-200MA-10KHZ-40KHZ | 20 kVp | 4 kW | 10kHz ... 40kHz |
| DF-HV20KV-200MA-10KHZ-50KHZ | 20 kVp | 4 kW | 10kHz ... 50kHz |
| DF-HV20KV-200MA-20KHZ | 20 kVp | 4 kW | 20 kHz |
| DF-HV20KV-200MA-20KHZ-30KHZ | 20 kVp | 4 kW | 20kHz ... 30kHz |
| DF-HV20KV-200MA-20KHZ-40KHZ | 20 kVp | 4 kW | 20kHz ... 40kHz |
| DF-HV20KV-200MA-20KHZ-50KHZ | 20 kVp | 4 kW | 20kHz ... 50kHz |
| DF-HV20KV-200MA-30KHZ | 20 kVp | 4 kW | 30 kHz |
| DF-HV20KV-200MA-30KHZ-40KHZ | 20 kVp | 4 kW | 30kHz ... 40kHz |
| DF-HV20KV-200MA-30KHZ-50KHZ | 20 kVp | 4 kW | 30kHz ... 50kHz |
| DF-HV20KV-200MA-40KHZ | 20 kVp | 4 kW | 40 kHz |
| DF-HV20KV-200MA-50KHZ | 20 kVp | 4 kW | 50 kHz |
| DF-HV20KV-200MA-50KHZ-100KHZ | 20 kVp | 4 kW | 50kHz ... 100kHz |
| DF-HV20KV-200MA-100KHZ | 20 kVp | 4 kW | 100 kHz |